Vernier Flash Photolysis Spectrometer (VSP-FP)

The current Vernier Flash Photolysis Spectrometer started shipping in April 2021 (serial number 63 and above). The complete reference manual, free experiments, and free software are available online at www.vernier.com/vsp-fp
The original Vernier Flash Photolysis Spectrometer was discontinued in March 2021 (serial number 62 and below). If you need its reference manual, you can find it here. It requires a previous version of free software, available here. A 600 nm interference filter is included with the original Vernier Flash Photolysis Spectrometer. A filter housing is also included. These filters select the wavelength monitored by the photodiode. Additional filters can be purchased from any filter company. The size that fits the housing is 12.5 mm (OD), 9.0 mm (ID).

Original Flash Photolysis Reference Manual

Free Flash Photolysis Software

TROUBLESHOOTING
-Primary Test:
1. Ensure the Flash Photolysis Spectrometer Software is installed on your computer before using the Vernier Flash Photolysis Spectrometer.
2. Connect the spectrometer to a powered USB port on the computer or a powered USB hub.
3. Start the Flash Photolysis Spectrometer Software on your computer.
4. Select the appropriate filter for your experiment.
5. Insert the filter housing into its slot on the instrument. The voltage reading should drop.

-Minimum software version: Free download available at www.vernier.com/vsp-fp
You do NOT need to purchase full LabView for this VI to work.
The Vernier Flash Photolysis Spectrometer Software collects the signal waveform from the photodetector. Subsequent data processing, such as conversion of the acquired voltage waveform into temporal profile of ΔA and fitting the ΔA profile to an appropriate function in order to extract the rate, can be exported as an ASCII CSV into Logger Pro or other third party software.

SPECIFICATIONS

General Specifications
Footprint (L x W) 4.25” x 5.5”
Weight 2 lbs
UL94 V-0
UL796 Meets requirements
USB cable 28 AWG shielded
Power source PC USB connection
Supplied current: 500 mA
Supplied voltage: 5 V
Altitude Up to 2500 m
Operating temperature 15 to 40 °C
Operating relative humidity 10 to 70% (non-condensing)
Spectrometer Specifications
Spectral coverage 450 to 750 nm
Spectral resolution Depends on the filter used. Typically, ~10 nm
Temporal resolution ~100 μs
Time window ≥ 15 ms
Detector Specifications
Detector type Silicon PIN
Active area 3.6 x 3.6 mm (13 mm2)
Wavelength range 350 to 1100 nm
Rise time 14 ns
Bias voltage 10 V
Dark current (with 1 MΩ load) 0.35 nA