Troubleshooting

  • Primary Test: Connect the Vernier Optical DO Probe to the data-collection interface. Start the data-collection software. Make sure you are using a current version of the software. The software will auto-ID the sensor.
  • Secondary Test: Verify the sensor reads 100% in air-saturated water or water-saturated air.

Note: The Optical DO Probe is not compatible with the Go!Link (GO-LINK).

GENERAL

CALIBRATION

The Optical DO Probe and sensor cap are factory calibrated and a calibration code specific to each individual cap is determined during the manufacturing process. The sensor is set to the stored calibration before shipping. If you wish to calibrate the sensor, you will conduct a one-point calibration using the saturated DO value. Refer to the user manual for detailed instructions.

OPERATION

EXPERIMENTS

MAINTENANCE and STORAGE

SPECIFICATIONS

  • Range
    • 0 to 20 mg/L
    • 0 to 300 %
  • Accuracy
    • +/- 0.2 mg/L below 10 mg/L
    • +/- 0.4 mg/L above 10 mg/L
    • +/- 2 % below 100%
    • +/- 5 % above 100%
  • Accuracy with calibration reset
    • +/- 0.1 mg/L below 10 mg/L
    • +/- 0.2 mg/L above 10 mg/L
    • +/- 1 % below 100%
    • +/- 5 % above 100%
  • Resolution: 0.006 mg/L (12 bit)
  • Response time: 90% of final reading in 40 seconds
  • Temperature compensation: automatic from 0 to 50℃
  • Pressure compensation: automatic from 30.4 kPa to 202.6 kPa (228 mmHg to 1519 mmHg)
  • Salinity compensation: manual, accounted for during calibration
  • Minimum sample flow: none