Vernier Mini GC (GC-MINI, discontinued)

- If you have the Go Direct Mini GC, see Go Direct Mini GC Troubleshooting and FAQs
- If you have the Mini GC Plus, see Mini GC Plus Gas Chromatograph Troubleshooting and FAQs
Troubleshooting
- Inconsistent Flow
⚬ Likely Cause: Worn or damaged septum.
⚬ Solution: Replace the septum. - Inconsistent Peak Areas
⚬ Likely Cause: Irregular injection volume or worn septum.
⚬ Solution: Check your injection technique and replace the septum. - No Communication
⚬ Likely Cause: Software mismatch or cable/Bluetooth connection.
⚬ Solution: Verify software versions and check USB cables. - No Power
⚬ Likely Cause: Power cord or switch issue.
⚬ Solution: Confirm that power plugs are fully inserted and the switch is turned on.
General Use Tips
- Use Approved Compounds Only: Always refer to the list of acceptable compounds in the user guide before injecting new samples.
- Avoid High-Water Samples: You CANNOT inject samples that contain more than 5% water. This will ruin the detector.
- Hamilton Syringe Note: The syringe shipped with the device has a brown bumper on the needle. Do not remove this bumper—it protects the injection port.
Also see Best Practice Techniques for Vernier gas chromatographs
Additional Troubleshooting
- Best Practice Techniques for Vernier gas chromatographs
- Can I inject water into Vernier gas chromatographs?
- Can I inject a gas into the Vernier gas chromatographs?
- What size needle do Vernier gas chromatographs require?
- Where can I order the syringe for Vernier gas chromatographs?
- The peak heights on my Vernier gas chromatograph are lower than they were previously.
- My Vernier gas chromatograph pressure won't get above 15 kPa.
- I need to ship a Vernier gas chromatograph to Vernier for repair. How should I package it?
- Can I purchase a replacement power supply for my Vernier gas chromatograph?
- Where can I buy the substances used in the Mini GC lab manual?
- My Mini GC or Mini GC Plus suddenly drops its connection.
- How long will the Mini GC/Mini GC Plus column last?
- Why does my Go!Link or GC Mini fail to connect every other time on Windows 8.1?
- How is the Go Direct Mini GC different than previous versions of the Mini GC?
Specifications
- Column: High-quality Restek MXT-1 Stainless capillary column
⚬ Length: 11 m
⚬ Inner diameter: 0.53 mm - Column operating temperature: 30°C to 120°C
⚬ maximum heating ramp: 10°C/min - Detector: Seacoast Science MEMS Chemi-Capacitive Detector
- Carrier gas: Ambient air or optional inert gas
- Operating pressure: 1 to 21 kPa (above ambient pressure)
- Environmental conditions
⚬ Temperature: 5°C to 40°C
⚬ Humidity: 95% or less (non-condensing environment) - Liquid injection volume: 0.01 to 1.0 µL
- Dimensions: 108 mm height, 191 mm length, 133 mm width
⚬ Weight: 1.3 kg - Power Requirements: 24 VDC, 2.5 A from an external power supply
- Utility: External power supply rated for 100–240 VAC, 1.5 A 50–60 Hz
Related Products
- Mini USB-C Cable (
CB-USB-C-MINI )
Replacement Parts
- GC Septa (package of 4) (
GC-SEP ) - Mini GC 1 μL Hamilton Syringe (
GC-SYR-MIC ) - Mini USB Cable (
CB-USB-MINI ) - Replacement Power Supply for Vernier Mini Gas Chromatographs (
GC2-MINI-PS )