Thermocouple (TCA-BTA )

Troubleshooting
- Primary Test: Connect the Thermocouple to an interface and check the accuracy of the readings for temperature in room air. They should be within 5 deg C.
- Secondary Test: Place your fingers over the wires and ensure that the temperature responds.
Additional Troubleshooting
- Can I purchase a replacement thermocouple wire for my TCA-BTA or GDX-TC sensor?
- How do you use the TCA-BTA to measure very low or very high temperatures more accurately?
- Can I use our thermocouple in a bunsen burner flame?
- Why doesn't my TCA-BTA work with Logger Pro or Logger Lite?
Specifications
- Temperature range: –200ºC to 1400ºC
- Typical Resolution: 0.40ºC
- Typical Accuracy:
⚬ –200ºC to 0ºC: ±5ºC
⚬ 0ºC to 900ºC: ±2.2º
⚬ 900ºC to 1400ºC: ±15ºC - Power: 11 mA @ 5 VDC
Calibration
Calibrate? No. This sensor is set to the stored calibration before shipping. If you wish to calibrate the sensor, you must conduct a three-point calibration. For information on the calibration coefficients used for this sensor, see Why doesn't my TCA-BTA work with Logger Pro or Logger Lite?
Related Products
- Wide-Range Temperature Probe (
WRT-BTA ) - Stainless Steel Temperature Probe (
TMP-BTA ) - Go!Temp (
GO-TEMP ) - EasyTemp (
EZ-TMP ) - Surface Temperature Sensor (
STS-BTA ) - Extra-Long Temperature Probe (
TPL-BTA ) - Go Direct® Thermocouple (
GDX-TC )