Troubleshooting

  • Primary Test: Attach the Rotary Motion Sensor to the Vernier data-collection interface and start the data-collection software. Start data collection, then Rotate the axle one revolution counterclockwise. The reading should advance from zero to approximately 6.28 radians.

Note: If you are using a LabPro interface, open the Rotary Motion Angular.cmbl file from the Probes and Sensors folder in Logger Pro.

Additional Troubleshooting

Specifications

  • Resolution: 1° or 0.25°
  • Optical Encoder: Bidirectional, quadrature encoder
  • Maximum Speed: 30 rev/s at 1° resolution 7.5 rev/s at 0.25° resolution
  • 3-step Pulley: 10 mm, 29 mm, and 48 mm groove diameter

Calibration

It is not necessary to calibrate this sensor.