Experiencing Bluetooth® issues with Go Direct® sensors on Chromebooks? Get help here.

Go Direct® Constant Current System User Manual

Order Code: GDX-CCS

The Go Direct Constant Current System is a DC power source with a built-in current probe designed for use in electrochemistry experiments. The user can set the current by tuning the dial. The system is capable of delivering up to 0.6 A. The following is a partial list of activities and experiments that can be performed using this sensor.

  • Electroplating and Faraday’s law
  • Electrolysis
  • Determining Avogadro’s number

Note: Vernier products are designed for educational use. Our products are not designed nor are they recommended for any industrial, medical, or commercial process such as life support, patient diagnosis, control of a manufacturing process, or industrial testing of any kind.

What's Included

  • Go Direct Constant Current System
  • AC power adapter
  • Micro USB Cable

Compatible Software

Choose a platform below to see its compatibility requirements.

LabQuest
InterfaceLabQuest App
LabQuest 3Full support
LabQuest 2 (discontinued)Full support 1
LabQuest (discontinued)Incompatible

Compatibility Notes

  1. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Computers
Software
InterfaceGraphical Analysis
No interface requiredFull support 1
LabQuest 3Full support 2
LabQuest 2 (discontinued)Full support 2 3

Compatibility Notes

  1. For Bluetooth® connectivity, only computers running Windows 10 or Mac OS X 10.10 or newer are supported. The computer must also have a compatible Bluetooth 4.0+ radio.
  2. Connect this sensor directly to a computer or Chromebook via a USB or wireless Bluetooth® connection for best results. Alternatively, this sensor is fully supported when used with LabQuest 2 or LabQuest 3 when the LabQuest is connected to a computer or Chromebook via Wireless Data Sharing.
  3. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Chromebook
Software
InterfaceGraphical Analysis
No interface requiredFull support
LabQuest 3Full support 1
LabQuest 2 (discontinued)Full support 1 2

Compatibility Notes

  1. Connect this sensor directly to a computer or Chromebook via a USB or wireless Bluetooth® connection for best results. Alternatively, this sensor is fully supported when used with LabQuest 2 or LabQuest 3 when the LabQuest is connected to a computer or Chromebook via Wireless Data Sharing.
  2. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
iOS
Software
InterfaceGraphical AnalysisGraphical Analysis GW
No interface requiredFull supportFull support
LabQuest 3Full support 1 2Full support 1 2
LabQuest 2 (discontinued)Full support 1 2 3Full support 1 2 3

Compatibility Notes

  1. Connect this sensor directly to a supported mobile device via a wireless Bluetooth® connection for best results.
  2. iOS and Androidâ„¢ devices can only connect to LabQuest 2 or LabQuest 3 via Wireless Data Sharing.
  3. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Android
Software
InterfaceGraphical AnalysisGraphical Analysis GW
No interface requiredFull supportIncompatible
LabQuest 3Full support 1 2Full support 1
LabQuest 2 (discontinued)Full support 1 2 3Full support 1 3

Compatibility Notes

  1. iOS and Androidâ„¢ devices can only connect to LabQuest 2 or LabQuest 3 via Wireless Data Sharing.
  2. Connect this sensor directly to a supported mobile device via a wireless Bluetooth® connection for best results.
  3. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Python
Software
InterfacePython
No interface requiredFull support
Javascript
Software
InterfaceJavascript
No interface requiredFull support 1

Compatibility Notes

  1. A wireless Bluetooth® connection can be made with Go Direct sensors on any platform running a browser with WebBluetooth support. However, only a few operating systems have been tested and testing has been limited.
LabVIEW
Software
InterfaceNI LabVIEW
No interface requiredFull support 1

Compatibility Notes

  1. For Bluetooth® connectivity, only computers running Windows 10 or Mac OS X 10.10 or newer are supported. The computer must also have a compatible Bluetooth 4.0+ radio.

Quick Start: Vernier Graphical Analysis® and Bluetooth®

  1. Turn on your sensor by plugging it into the AC power adapter. The Bluetooth® LED will blink red.
  2. Launch Graphical Analysis, then click Sensor Data Collection.
  3. Select your sensor from the list. The sensor ID is located on the sensor label near the bar code. Note: If you don’t see a list of available sensors, click WIRELESS. After selecting your sensor, click Pair.
  4. Click DONE. You are now ready to collect data.

Using other Vernier data-collection apps or want to connect via USB?

Visit www.vernier.com/start-go-direct

Note: This sensor also works with LabQuest 2 and LabQuest 3; it does not work with the original LabQuest.

Connecting the Sensor

See the following link for up-to-date connection information:

www.vernier.com/start/gdx-ccs

Connecting via Bluetooth

Ready to connect Red LED next to Bluetooth icon flashes when the sensor is awake and ready to connect. At the same time, the red LED next to the power icon is steady.
Connected Green LED next to Bluetooth icon flashes when the sensor is connected via Bluetooth. At the same time, the red LED next to the power icon is on steady.

Connecting via USB

Connected Red LED next to power icon on steady when sensor is connected to Graphical Analysis via USB. LED next to Bluetooth icon off.

Powering the Sensor

Turning on the sensor

Red LED is on steady when connected to AC power.

Putting the sensor in sleep mode

Red LED is off when disconnected from AC power.

Specifications

Range

0 to 0.6 A

Linearity

0.1%

Supply Voltage

5 VDC

Supply current (typical)

9 mA

Output voltage range

0 to 5 V

Transfer function

Vout=6.9(I) + 0

Power supply (input)

  • 100–240 Volts AC
  • 50/60 Hz 0.5 A
  • 26–41 VA

Power supply (output)

5 V, 1.5 A

USB specification

2.0

Wireless specification

Bluetooth 4.2

Maximum wireless range

30 m

Dimensions

9 cm length, 6 cm height, 5 cm depth, leads: 60 cm long

How the Sensor Works

The Go Direct Constant Current System is a DC-coupled amplifier capable of delivering up to 0.6 A at 5 V DC. A current-sensing resistor allows monitoring of the output current by the data-collection software.

Current Conventions

The current will be indicated as positive if current flows from the positive (red) terminal to the negative (black) terminal as described by conventional current flow notation. Using electron flow notation, electrons flow out of the negative terminal through the circuit back to the positive terminal. The negative terminal is connected to ground via a power transistor and 0.1 Ω sensing resistor. The positive terminal connects to a 5 V power supply. The Constant Current System dial can be tuned to deliver up to 0.6 A with the voltage adjusting automatically (max 5 V). If the electrolyte resistance is too high, the 0.6 A may not be attained before the maximum of 5 V is reached.

Troubleshooting

For troubleshooting and FAQs, see www.vernier.com/til/3851

Repair Information

If you have followed the troubleshooting steps and are still having trouble with your Go Direct Constant Current System, contact Vernier Technical Support at support@vernier.com or call 888-837-6437. Support specialists will work with you to determine if the unit needs to be sent in for repair. At that time, a Return Merchandise Authorization (RMA) number will be issued and instructions will be communicated on how to return the unit for repair.

Accessories/Replacements

Item Order Code
LabQuest Power Supply

LQ3-PS

Micro USB Cable

CB-USB-MICRO

USB-C to Micro USB Cable

CB-USB-C-MICRO

Warranty

Warranty information for this product can be found on the Support tab at www.vernier.com/gdx-ccs/#support

General warranty information can be found at www.vernier.com/warranty

Disposal

When disposing of this electronic product, do not treat it as household waste. Its disposal is subject to regulations that vary by country and region. This item should be given to an applicable collection point for the recycling of electrical and electronic equipment. By ensuring that this product is disposed of correctly, you help prevent potential negative consequences on human health or on the environment. The recycling of materials will help to conserve natural resources. For more detailed information about recycling this product, contact your local city office or your disposal service.

Battery recycling information is available at www.call2recycle.org

Do not puncture or expose the battery to excessive heat or flame.

The symbol, shown here, indicates that this product must not be disposed of in a standard waste container.

Contact Support

Fill out our online support form or call us toll-free at 1-888-837-6437.

SAVE/SHARE YOUR CART