Electrode Amplifier User Manual
Order Code: EA-BTA
Use our Electrode Amplifier to collect data with almost any electrode that has a BNC connector. This includes electrodes from Vernier such as one of our pH, ISE, or ORP electrodes, or third-party electrodes, such as pH, ISE (ion-selective electrode), or ORP (oxidation-reduction potential). The Electrode Amplifier can produce an output in mV, pH (when using a pH electrode). You can also perform your own 2-point linear calibration.
Note: Vernier products are designed for educational use. Our products are not designed nor are they recommended for any industrial, medical, or commercial process such as life support, patient diagnosis, control of a manufacturing process, or industrial testing of any kind.
Compatible Software
Choose a platform below to see its compatibility requirements.LabQuest
Interface LabQuest App LabQuest 3 Full support 1 LabQuest 2 (discontinued) Full support 1 LabQuest (discontinued) Full support 1 Compatibility Notes
Computers
Software Interface Graphical Analysis Logger Pro Logger Lite LabQuest Mini Full support 1 Full support 1 Full support 1 LabQuest 3 Full support 1 Full support 1 Incompatible 1 LabQuest 2 (discontinued) Full support 1 Full support 1 Full support 1 LabQuest Stream (discontinued) Full support 1 2 Full support 1 Full support 1 2 Go!Link Full support 1 Full support 1 Full support 1 LabQuest (discontinued) Full support 1 Full support 1 Full support 1 LabPro (discontinued) Incompatible Full support 1 Full support 1 Compatibility Notes
Chromebook
Software Interface Graphical Analysis LabQuest Mini Full support 1 LabQuest 3 Full support 1 LabQuest 2 (discontinued) Full support 1 LabQuest Stream (discontinued) Full support 1 2 Go!Link Full support 1 LabQuest (discontinued) Full support 1 Compatibility Notes
iOS
Software Interface Graphical Analysis Graphical Analysis GW LabQuest Stream (discontinued) Full support 1 Partial support 1 2 LabQuest 3 Full support 1 3 Full support 1 3 LabQuest 2 (discontinued) Full support 1 3 Full support 1 3 Compatibility Notes
Android
Software Interface Graphical Analysis Graphical Analysis GW LabQuest Stream (discontinued) Full support 1 Incompatible 1 LabQuest 3 Full support 1 2 Full support 1 2 LabQuest 2 (discontinued) Full support 1 2 Full support 1 2 Compatibility Notes
Arduino
Software Interface Arduino Vernier Arduino® Interface Shield Full support LabVIEW
Software Interface NI LabVIEW SensorDAQ (discontinued) Full support Vernier myDAQ Adapter Full support 1 Go!Link Full support LabQuest Mini Full support LabQuest Stream (discontinued) Full support LabQuest 3 Full support LabQuest 2 (discontinued) Full support LabQuest (discontinued) Full support Compatibility Notes
Texas Instruments
Software Interface EasyData DataMate TI-84 SmartView DataQuest TI-Nspire Software EasyLink Full support 1 2 Incompatible Full support 1 3 Full support 1 Full support 1 3 CBL 2 (discontinued) Full support 1 4 Full support 1 4 5 Incompatible Incompatible Incompatible LabPro (discontinued) Full support 1 4 Full support 1 4 5 Incompatible Incompatible Incompatible TI-Nspire Lab Cradle (discontinued) Incompatible Incompatible Incompatible Full support 1 Full support 1 Compatibility Notes
Getting Started
- Connect the sensor to the interface (LabQuest Mini, LabQuest 3, etc.).
- Start the appropriate data-collection software (Vernier Graphical Analysis®, LabQuest® App, or Logger Proâ„¢) if not already running, and choose New from File menu. The software will identify the sensor and load a default data‑collection setup. You are now ready to collect data.
If you are collecting data using a Chromebook™, mobile device such as iPad® or Android™ tablet, or a Vernier wireless interface, please see the following link for up-to-date connection information:
Specifications
Power |
7 mA @ 5VDC |
Input range |
–450 mV to +1100 mV |
Impedance |
100 M Ohm |
Gain |
2.2 V/V |
Offset |
1.20 V (nominal) |
Stored calibration (mV) |
|
Stored calibration (pH) |
|
Care and Maintenance
Do not wrap the cable tightly around the sensor for storage. Repeatedly doing so can irreparably damage the wires and is not covered under warranty.
How the Sensor Works
The Electrode Amplifier is a circuit that allows the millivolt output of an electrode to be monitored by an interface. The electrode is connected to the BNC connector on one end of the amplifier. The cable from the Electrode Amplifier connects to the interface.
The Electrode Amplifier will report values in units of pH. If an ORP or ISE electrode is attached, however, you can choose to read in units of mV. Custom linear 2-point calibrations can also be performed if desired.
Troubleshooting
For troubleshooting and FAQs, see www.vernier.com/til/1430
Repair Information
If you have watched the related product video(s), followed the troubleshooting steps, and are still having trouble with your Electrode Amplifier, contact Vernier Technical Support at support@vernier.com or call 888-837-6437. Support specialists will work with you to determine if the unit needs to be sent in for repair. At that time, a Return Merchandise Authorization (RMA) number will be issued and instructions will be communicated on how to return the unit for repair.
Accessories/Replacements
Item | Order Code |
---|---|
NH4-BNC |
|
CA-BNC |
|
CL-BNC |
|
FPH-BNC |
|
GPH-BNC |
|
NO3-BNC |
|
ORP-BNC |
|
PH-BNC |
|
K-BNC |
Warranty
Warranty information for this product can be found on the Support tab at www.vernier.com/ea-bta/#support
General warranty information can be found at www.vernier.com/warranty
Disposal
When disposing of this electronic product, do not treat it as household waste. Its disposal is subject to regulations that vary by country and region. This item should be given to an applicable collection point for the recycling of electrical and electronic equipment. By ensuring that this product is disposed of correctly, you help prevent potential negative consequences on human health or on the environment. The recycling of materials will help to conserve natural resources. For more detailed information about recycling this product, contact your local city office or your disposal service.
The symbol, shown here, indicates that this product must not be disposed of in a standard waste container.
Contact Support