Experiencing Bluetooth® issues with Go Direct® sensors on Chromebooks? Get help here.

Go Direct® Gas Pressure Sensor User Manual

Order Code: GDX-GP

Go Direct Gas Pressure is used to monitor pressure changes in gas-law experiments in chemistry, biology, and physics.

Note: Vernier products are designed for educational use. Our products are not designed nor are they recommended for any industrial, medical, or commercial process such as life support, patient diagnosis, control of a manufacturing process, or industrial testing of any kind.

What's Included

  • Go Direct Gas Pressure
  • Two tapered valve connectors inserted into a No. 5 stopper
  • One tapered valve connector inserted into a No. 1 stopper
  • One two-way valve
  • Two Luer-lock connectors connected to either end of a piece of plastic tubing
  • One 20 mL syringe
  • Two transpiration tubing clamps
  • Micro USB Cable

Compatible Software

Choose a platform below to see its compatibility requirements.

LabQuest
InterfaceLabQuest App
LabQuest 3Full support
LabQuest 2 (discontinued)Full support 1
LabQuest (discontinued)Incompatible

Compatibility Notes

  1. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Computers
Software
InterfaceGraphical Analysis
No interface requiredFull support 1
LabQuest 3Full support 2
LabQuest 2 (discontinued)Full support 2 3

Compatibility Notes

  1. For Bluetooth® connectivity, only computers running Windows 10 or Mac OS X 10.10 or newer are supported. The computer must also have a compatible Bluetooth 4.0+ radio.
  2. Connect this sensor directly to a computer or Chromebook via a USB or wireless Bluetooth® connection for best results. Alternatively, this sensor is fully supported when used with LabQuest 2 or LabQuest 3 when the LabQuest is connected to a computer or Chromebook via Wireless Data Sharing.
  3. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Chromebook
Software
InterfaceGraphical Analysis
No interface requiredFull support
LabQuest 3Full support 1
LabQuest 2 (discontinued)Full support 1 2

Compatibility Notes

  1. Connect this sensor directly to a computer or Chromebook via a USB or wireless Bluetooth® connection for best results. Alternatively, this sensor is fully supported when used with LabQuest 2 or LabQuest 3 when the LabQuest is connected to a computer or Chromebook via Wireless Data Sharing.
  2. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
iOS
Software
InterfaceGraphical AnalysisGraphical Analysis GW
No interface requiredFull supportFull support
LabQuest 3Full support 1 2Full support 1 2
LabQuest 2 (discontinued)Full support 1 2 3Full support 1 2 3

Compatibility Notes

  1. Connect this sensor directly to a supported mobile device via a wireless Bluetooth® connection for best results.
  2. iOS and Androidâ„¢ devices can only connect to LabQuest 2 or LabQuest 3 via Wireless Data Sharing.
  3. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Android
Software
InterfaceGraphical AnalysisGraphical Analysis GW
No interface requiredFull supportIncompatible
LabQuest 3Full support 1 2Full support 1
LabQuest 2 (discontinued)Full support 1 2 3Full support 1 3

Compatibility Notes

  1. iOS and Androidâ„¢ devices can only connect to LabQuest 2 or LabQuest 3 via Wireless Data Sharing.
  2. Connect this sensor directly to a supported mobile device via a wireless Bluetooth® connection for best results.
  3. Wireless connection of this sensor to a LabQuest 2 requires a Go Direct Bluetooth Adapter when used with some older LabQuest 2 models.
Python
Software
InterfacePython
No interface requiredFull support
Javascript
Software
InterfaceJavascript
No interface requiredFull support 1

Compatibility Notes

  1. A wireless Bluetooth® connection can be made with Go Direct sensors on any platform running a browser with WebBluetooth support. However, only a few operating systems have been tested and testing has been limited.
LabVIEW
Software
InterfaceNI LabVIEW
No interface requiredFull support 1

Compatibility Notes

  1. For Bluetooth® connectivity, only computers running Windows 10 or Mac OS X 10.10 or newer are supported. The computer must also have a compatible Bluetooth 4.0+ radio.

Quick Start: Vernier Graphical Analysis® and Bluetooth®

  1. Charge your sensor for at least 2 hours before first use.
  2. Turn on your sensor. The LED will blink red.
  3. Launch Graphical Analysis, then click Sensor Data Collection.
  4. Select your sensor from the list. The sensor ID is located on the sensor label near the bar code. Note: If you don’t see a list of available sensors, click WIRELESS. After selecting your sensor, click Pair.
  5. Click DONE. You are now ready to collect data.

Using other Vernier data-collection apps or want to connect via USB?

Visit www.vernier.com/start-go-direct

Note: This sensor also works with LabQuest 2 and LabQuest 3; it does not work with the original LabQuest.

Charging the Sensor

Connect Go Direct Gas Pressure to the included Micro USB Cable and any USB device for two hours.

You can also charge up to eight Go Direct Gas Pressure Sensors using our Go Direct Charge Station, sold separately (order code: GDX-CRG). An LED on each Go Direct Gas Pressure indicates charging status.

Charging

Orange LED next to the battery icon is solid while the sensor is charging.

Fully charged

Green LED next to the battery icon is solid when the sensor is fully charged.

Powering the Sensor

Turning on the sensor

Press button once. Red LED indicator flashes when unit is on.

Putting the sensor in sleep mode

Press and hold button for more than three seconds to put into sleep mode. Red LED indicator stops flashing when sleeping.

Connecting the Sensor

See the following link for up-to-date connection information:

www.vernier.com/start/gdx-gp

Connecting via Bluetooth

Ready to connect Red LED next to the Bluetooth icon flashes when sensor is awake and ready to connect.
Connected Green LED next to the Bluetooth icon flashes when sensor is connected via Bluetooth.

Connecting via USB

Connected and charging Orange LED next to the battery icon is solid when the sensor is connected to Graphical Analysis via USB and the unit is charging. LED next to Bluetooth icon is off.
Connected, fully charged Green LED next to the battery icon is solid when the sensor is connected to Graphical Analysis via USB and fully charged. LED next to Bluetooth icon is off.
Charging via USB,
connected via Bluetooth
Orange LED next to the battery icon is solid when the sensor is charging. Green LED next to the Bluetooth icon flashes.

Identifying the Sensor

When two or more sensors are connected, the sensors can be identified by tapping or clicking Identify in Sensor Information.

Using the Product

Connect the sensor following the steps in the Quick Start section of this user manual.

Important: The Go Direct Gas Pressure sensing element will be damaged with direct contact to liquid.

Videos

Calibrating the Sensor

You do not have to perform a new calibration when using Go Direct Gas Pressure. The sensor is calibrated prior to shipping.

If you do choose to calibrate, a one-point calibration at atmospheric pressure is adequate for most applications. Go Direct Gas Pressure has been calibrated to read station pressure. Station pressure is the true atmospheric pressure at your location, or station. If you prefer it to read sea-level pressure for conducting weather studies, you can perform a one-point calibration to correct for elevation, using the sea-level corrected pressure value for your location obtained from a reputable source (e.g., NOAA, Weather Underground, etc.). Sea-level pressure is the pressure after the station pressure has been adjusted to its equivalent pressure at sea level. This is commonly done to normalize pressures at various altitudes for weather forecasts.

To calibrate the sensor in Graphical Analysis, complete the following steps.

  1. Click or tap the sensor meter to view sensor options.
  2. Select Calibrate and follow the applicable prompts on the Calibrate Sensor screen.

Once you have calibrated a Go Direct sensor, the calibration is automatically stored to the sensor and will be used each time you connect to your device. You can always choose to restore factory defaults if you feel the custom calibration is invalid.

Specifications

Pressure range

0 kPa to 400 kPa

Maximum pressure without permanent damage

410 kPa

Typical accuracy

±3 kPa

Internal volume

0.8 mL

Resolution

0.03 kPa

Maximum sampling rate

50 samples/s

USB specification

2.0

Wireless specification

Bluetooth 4.2

Maximum wireless range

30 m

Dimensions

9.5 cm × 6.0 cm × 3.25 cm

Battery

300 mA Li-Poly

Battery life (single full charge)

~24 hours with active use

Battery life (long term)

~500 full charge cycles (several years depending on usage)

Care and Maintenance

Battery Information

Go Direct Gas Pressure contains a small lithium-ion battery. The system is designed to consume very little power and not put heavy demands on the battery. Although the battery is warranted for one year, the expected battery life should be several years. Replacement batteries are available from Vernier (order code: GDX‑BAT-300).

Storage and Maintenance

To store Go Direct Gas Pressure for extended periods of time, put the device in sleep mode by holding the Power button down for at least three seconds. The red LED will stop flashing to show that the unit is in sleep mode. Over several months, the battery will discharge but will not be damaged. After such storage, charge the device for a few hours, and the unit will be ready to go.

Exposing the battery to temperatures over 35°C (95°F) will reduce its lifespan. If possible, store the device in an area that is not exposed to temperature extremes.

Water Resistance

Go Direct Gas Pressure is not water resistant and should never be immersed in water.

If water gets into the device, immediately power the unit down (press and hold the power button for more than three seconds). Disconnect the sensor and charging cable, and remove the battery. Allow the device to dry thoroughly before attempting to use the device again. Do not attempt to dry using an external heat source.

How the Sensor Works

The sensor in this unit has a membrane that flexes as pressure changes. This sensor is arranged to measure absolute pressure. One side of the membrane is a vacuum, while the other side is open to the atmosphere. The sensor produces an output voltage that varies in a linear way with absolute pressure. It includes special circuitry to minimize errors caused by changes in temperature.

Troubleshooting

For troubleshooting and FAQs, see www.vernier.com/til/3850

Repair Information

If you have watched the related product video(s), followed the troubleshooting steps, and are still having trouble with your Go Direct Gas Pressure, contact Vernier Technical Support at support@vernier.com or call 888-837-6437. Support specialists will work with you to determine if the unit needs to be sent in for repair. At that time, a Return Merchandise Authorization (RMA) number will be issued and instructions will be communicated on how to return the unit for repair.

Accessories/Replacements

Warranty

Warranty information for this product can be found on the Support tab at www.vernier.com/gdx-gp/#support

General warranty information can be found at www.vernier.com/warranty

Contact Support

Fill out our online support form or call us toll-free at 1-888-837-6437.

SAVE/SHARE YOUR CART