...Equation: power function Y = C*X^p C: -2.995 p: 0.9054 RELATED VERNIER PRODUCTS Ethanol Cap Assemblies (3) (ETH-CAPS) #6 Stopper for Ethanol Sensor (ETH-STOP) Tape for Ethanol Sensor (ETH-TAPE) 250......
...in Investigating Biology through Inquiry (BIO-I) You can find these labs at: https://www.vernier.com/products/sensors/eth-bta/ The Go Direct® Ethanol Vapor (GDX-ETOH) or the Ethanol Sensor (ETH-BTA) can be used for these experiments.......
...measuring an intermediate range you can also calibrate using 0.2% and 2.0% ethanol. Both the Go Direct® Ethanol Vapor (GDX-ETOH) and the Ethanol Sensor (ETH-BTA) can be calibrated this way.......
The Ethanol Sensor (ETH-BTA) is not compatible with Go!Link (GO-LINK). When connected together, Logger Pro and Logger Lite will incorrectly identify the Ethanol Sensor as an O2 Gas sensor. The......
No. The Go Direct® Ethanol Vapor (GDX-ETOH) and Ethanol Sensor (ETH-BTA) are gas sensors that detect ethanol vapor. The sensors needs to be placed above a liquid sample. The use......
The Go Direct® Ethanol Vapor (GDX-ETOH) and Ethanol Sensor (ETH-BTA) should not be exposed to concentrations of ethanol greater than 3%. To determine the amount of ethanol produced from a......
If you want to use a Vernier CO2 Gas Sensor and a Vernier ethanol sensor to show that CO2 gas and Ethanol are both produced during fermentation, you should use......
Vernier ethanol sensors work well between 0% and 3% ethanol. The standard calibration will give good change over time data, but for quantitative data, calibration will increase the accuracy of......
All Vernier ethanol sensors need to warm up for 5 minutes before use. On first use, the ethanol sensor may need more than 5 minutes to warm up. If the......
All Vernier ethanol sensors will never give a reading of in air because of the type of calibration formula that is used. The minimum reading will typically fall within the......