Troubleshooting

  • Primary Test: Are you using the UVB Sensor to check UVA light? Confirm that the sensor is appropriate for the experiment.
  • Secondary Test: Take readings outdoors in the sunlight with and without a barrier, such as a sunglasses lens. Confirm that the readings are in the normal range for your area.

Additional Troubleshooting

Specifications

  • UV peak sensitivity: one Volt per 204 mW/m2 at 300 nm
  • Wavelength sensitivity region, approximate: 265 to 322 nm, 10% sensitivity points.
  • Typical Resolution: 0.25 mW/m2
  • Dimensions: 21 cm by 2 cm diameter
  • Time response: ~ 2 seconds to reach 95% of final reading
  • Stored calibration coefficients: slope (gain) = 204 mW/(m² V); intercept (offset) = 0

Calibration

Calibrate? No. The sensor is set to the stored calibration before shipping. It is quite difficult to calibrate a UV sensor to read in absolute units, since you must have a source of known UV intensity and spectral distribution. More often, you will simply want to calibrate the sensor in terms of a relative intensity as noted below.

  1. Cover the tip of the UVB Sensor with a clean opaque object.
  2. Select the calibration option of the program you are using.
  3. Enter 0 (zero) as the first known intensity, then click or tap Keep.
  4. Now allow full UV intensity to strike the sensor.
    1. Since the orientation of the sensor affects the reading, it is best to hold the sensor in place with a ring stand or other clamp.
    2. To point the sensor directly at the sun, make the shadow of the sensor tube as small as possible.
  5. Enter 100 as the next known intensity, then click or tap Keep.
  6. Exit the calibration dialog. Measurements of the sensor will be relative to this second intensity.

For more information on calibration, see How do I calibrate my sensor?

Related Products